BMC Y Series BPAP System

Brand
product img

Introduction: 

 

BMC Y series BPAP Systems are Bi-level PAP (Bi-level Positive Airway Pressure) devices intended to
provide non-invasive ventilation for patients with Obstructive Sleep Apnea (OSA) and Respiratory
Insufficiency. They are intended for adult patients by prescription in the home or hospital/institutional
environment. With its Target Tidal Volume function and other excellent comfort features and effective
performance, it offers each patient personalized ventilation support.

Main Expected Benefits for Patients of Home Non-invasive Ventilation in COPD: An International Survey of Indications
and Practices from 15 Countries. (Order of importance)

  • Hospital readmission reduction
  • Quality of life improvement
  • Dyspnea relief
  • Gas exchange improvement
  • Exacerbations reduction
  • Sleep quality improvement
  • Exercise tolerance
  • Survival improvement FEV1 improvement

 

Comfortability:

Humidifier:

  • Eco Smart heating system with innovative dual water chambers design delivers accurate water quantity control
  • with real time compensation, ensuring excellent humidifying capacity and improves comfort.
  • Easy to take off and clean.
  • Function to prevent overheating when water is out.

 

Inspiratory / Expiratory sensitivity:

8 grades of I Sens and E Sens optimize the compliance of machine with the patient.

 

Efficiency In Therapy:

S/T Mode
Machine complies with patient breathing. However, if there is no inhalation for a certain period of time, the machine
will give a forced ventilation to ensure the minimum ventilation.

 

Target Tidal Volume Function
Optimize IPAP according to mean Vt of last 5 breathings and prescribed Target Vt. Larger difference between mean Vt and
target Vt takes more evident adjustment in IPAP. Inspiratory pressure is between IPAP min and IPAP max. Larger difference between mean Vt and target Vt takes more evident adjustment in IPAP. Patient Vt is the mean of Vt values from patient’s last 5 breathings.

 

Auto Leak Compensation
The machine detects the leaks during treatment in real time and adjusts the baseline to ensure correct triggering and related  functions.

Inspiratory time control
Tim min and Ti max could be set independently, avoiding insufficient ventilation due to short inspiratory time. At the  meantime, cases can be prevented where expiratory sensitivity is unable to meet due to large leaks.

 

Specification 

Model Comparison

Y-25T IPAP: 4 - 25 hPa
EPAP: 4 - 25 hPa
CPAP mode: 4 - 20 hPa
3.5-inch CPAP, S, T, S/T
Y-30T IPAP: 4 - 30 hPa
EPAP: 4 - 25 hPa
CPAP mode: 4 - 20 hPa
3.5-inch CPAP, S, T, S/T

 

General Info
Dimensions: 170 mm × 180 mm × 118 mm
290 mm × 180 mm × 134 mm (with the humidifier)
Weight: 1.5 kg
2.5 kg (with the humidifier)
Water capacity: 350 mL at recommendedwater level
Ramp
The ramp time ranges from 0 to 60 minutes
Humidifier
Humidifier Settings: off, 1 to 5 (95°F to 167°F
/ 35°C to 75°C)
Humidifier Output: No less than 10 mg H2O/L
SpO2
Range: 0 to 100%
Pulse Rate
Range: 40 to 240 BPM
Sound Pressure Level
< 30 dB, when the device is working at the
pressure of 10 hPa
Storage
SD card can record patient data and fault
information
AC Power Consumption
100 - 240 V AC, 50/60 Hz, Max 2 A


Key Parameters
Target Vt: On/ Off
150~1500 mL
Reslex: Patient, Off, 1~3
I Sens.: 1~8
E Sens.: 1~8
Res Rate: 3~40 BPM
Ti: 0.3-3.0s
Rise Time: 1~4